Tighten the baseplate clamp snugly.
Tighten the flexible coupler's clamp onto the drive shaft.
Tighten the clamp of the flexible coupler onto the drive shaft of the Z-drive unit using the 5/64
inch Allen wrench.
Note: The flexible coupler clamp must be securely tightened or the drive may slip.
Lightly tighten the vertical and horizontal adjustment screws as shown. Since the motor is
usually self-aligning, this should be a good position for the drive. Check the alignment by noting
the drag while rotating the right-hand fine focus knob. No noticeable drag should be felt for the
full 360º rotation of the fine focus knob. If any drag is felt other than the slight drag of the
gears, loosen the vertical and horizontal adjustment screws and move the drive to a point
where no drag is felt. Then tighten the vertical and horizontal adjustment screws.
Tighten the vertical adjustment screws after alignment.