Olympus MX61L Panduan Perakitan - Halaman 39

Jelajahi secara online atau unduh pdf Panduan Perakitan untuk Mikroskop Olympus MX61L. Olympus MX61L 49 halaman. 200mm/300mm compatible semiconductor/fpd inspection microscopes

Olympus MX61L Panduan Perakitan

Glare Prevention Setting

Functions
1. Sets the glare prevention.
2. Cancels the glare prevention.
Special notes
1. The glare prevention is set or canceled according to the setup of DIP switch No. 4 at the moment the microscope is
turned ON. This command is used to change this initial condition.
2. The glare prevention setting of the MX61/61L made using instruction command 1FPV
switch setting. The setting using 1FPV
mode, until the microscope is turned OFF.
Command descriptions
Command
Type
1FPV
ON
Request
1FPV
OFF
Request
1FPV?
Query
)
"
" represents a space.
*
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
(See page 20.)
ON is enabled even after the microscope is switched the remote mode to local
Description
Setting the anti-glare setting.
(in darkfield observation only)
Cancellation of the anti-glare
setting.
Check of the current glare
prevention setting status.
ON is given priority over the DIP
Response
1FPV
+
Setting is complete.
1FPV
!,E01120
A string other than ON/OFF is
sent after "1FPV
1FPV
+
Cancellation is complete.
1FPV
!,E01120
A string other than ON/OFF is
sent after "1FPV
1FPV
ON
Glare prevention is ON.
1FPV
OFF
Glare prevention is OFF.
1FPV
!,E01150
A string is sent after "1FPV?
MX61/MX61L
1FPV 1FPV?
Description
".
".
".