Olympus Pom Petunjuk Manual - Halaman 13

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus Pom. Olympus Pom 15 halaman.

Olympus Pom Petunjuk Manual
life
BODY
Test
Plates:
Insert the 1/4 wave length retardation plate Imica 147m,1'l or the
ting plate Igypsum 530m/1) into the provided windw at an angle of 45°.
They are used for examiation of the bi-refringence or warps and for
determination of the axis.
+
SPECIAL ACCESSORIES:
Berek Compensator:
Angle Indicator Dial.
A piece of calcite is set in this compensator to measure the
bi-refringence of the specimen.
The compensator is inserted into the
test plate window.
Readings taken
from
the angle indicator dial
combined with the data taken from the attached compensation curve
table will determine the retardation.
A bi-refringent specimen may be set at the extinct position stage
45° by the scale and clamp it at the diagonal posittion.
Then insert
the compensator, turn the dial until the
interferential color in zero
order appears in the center of the field.
Take the reading of the
angle.
Compare it on the compensation curve and obtain
the micron
valus of the retardation.
If the order doesn't lessen even by turning
the compensation dial, turn
the specimen 90° and repeat the same
procedure.
At the position of 30° on the indicator dial, the tcst plate will
become perpendicular to the optical axis with 0 retardation; therefore,
insertion and removal of the test plate is performed at this position.
If it is overly slanted it will get stuck at the inserting window.
Eyepiece Adaptor.
If the observer prefers to use an evepiece of
a standard microscope [with a smaller .lIS diamet-
en, he may remove the
['O\'[
eyepiece and replace
with this adaptor.
Then the normal evepiece may
be inserted in it.
This will make possible the
of the general eyepieces,
grain-size
measuring
evepieces, other specialized eyepieces.