Olympus CKX41 Petunjuk Manual - Halaman 29
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Peralatan Laboratorium Olympus CKX41. Olympus CKX41 36 halaman. Reflected fluorescence system
Juga untuk Olympus CKX41: Ikhtisar (7 halaman), Brosur & Spesifikasi (4 halaman), Manual Perbaikan (43 halaman)
- 1. Table of Contents
- 1. General Precautions for Observation
- 1. Using the Controls
- 2. Important
- 2. Applicable Fluorescence Objectives
- 3. Nomenclature
- 3. Centering the Mercury Burner
- 4. Troubleshooting Guide
- 4. Specifications
- 4. Assembly
- 4. Assembly
- 4. U-Excitation Filter Set Ckx-Nu (Optional)
- 4. Controls of each Module
- 4. Switching the Filter Slider
- 5. Proper Selection of the Power Supply Cord
- 5. Summary of Reflected Fluorescence Observation Procedure
Fig. 19
6
Fig. 20
5. Set the burner type switch 5 on the rear of the power
supply unit according to the type of the mounted burner
(L1 or L2). (Fig. 19)
# If the burner socket clamping screw 1 is loosened while
the burner is on, the interlock switch turns off the burner.
To restart the burner, first set the main switch of the power
5
supply unit to "
Then, after firmly tightening the clamping screw, set the
main switch to " " (ON) again.
* The burner cannot be re-ignited for about 10 minutes,
that is, until the mercury vapor inside it has cooled down
and condenser to liquid.
Caution for burner replacement during observation
The burner, lamp housing and its surroundings are very
hot during operation.
Before replacing the burner, set the main switch to "
(OFF), unplug the power cord and wait until the burner,
lamp housing and its surroundings cool down completely.
6. Press the reset button 6 on the front of the power supply
unit to reset the hour counter display to "0.0". (Fig. 20)
}The unit of hour counter display is hour. Replace the burner
when the displayed time reaches "100.0" hours.
Note that the above-mentioned life of a mercury burner may
expire before the above-mentioned average depending on
how frequently the main switch is set to " " (ON) and " "
(OFF). When ignition of the burner deteriorates or flickering
becomes noticeable, replace the burner regardless of the
hour counter display.
# Be sure to reset to "0.0". Otherwise, the burner may not
be ignitable.
Clear mode than 10 cm of space around the power
supply unit to assure proper heat radiation.
The power supply unit should be installed so that the
main switch can easily be accessed and set to "
in case of emergency.