Olympus CKX41 Petunjuk Manual - Halaman 31
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Peralatan Laboratorium Olympus CKX41. Olympus CKX41 36 halaman. Reflected fluorescence system
Juga untuk Olympus CKX41: Ikhtisar (7 halaman), Brosur & Spesifikasi (4 halaman), Manual Perbaikan (43 halaman)
- 1. Table of Contents
- 1. General Precautions for Observation
- 1. Using the Controls
- 2. Important
- 2. Applicable Fluorescence Objectives
- 3. Nomenclature
- 3. Centering the Mercury Burner
- 4. Troubleshooting Guide
- 4. Specifications
- 4. Assembly
- 4. Assembly
- 4. U-Excitation Filter Set Ckx-Nu (Optional)
- 4. Controls of each Module
- 4. Switching the Filter Slider
- 5. Proper Selection of the Power Supply Cord
- 5. Summary of Reflected Fluorescence Observation Procedure
9
Fig. 24
REFLECTED FLUORESCENCE SYSTEM
Fuse replacement (Fig. 24)
# Set the main switch to "
cord before replacing fuses.
1. Remove each of the fuse holders 9 by turning it counter-
clockwise using a flat-blade screwdriver and pulling out.
2. Replace both fuses with new ones.
# Always use the designated fuses. Otherwise, a fire hazard
may result.
Applicable fuses
" (OFF) and unplug the power
U-RFTL50-100:
T4A(H)250V
(LITTELFUSE 215004), 2 pieces
U-RFTL50-200:
T2.5A(H)250V
(LITTELFUSE 215025), 2 pieces
FOR CKX41
28