AccuMac AM1660 Panduan Pengguna - Halaman 5

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Termometer AccuMac AM1660. AccuMac AM1660 10 halaman. Precision industrial platinum resistance thermometer

AccuMac AM1660 Panduan Pengguna
AM1660 Precision Industrial PRT User's Guide

2 Specifications

2.1 Specifications
Temperature Range
Resistance at 0 °C
Temperature Coefficient
Accuracy
Drift
Short Term Stability
Thermal Shock
Hysteresis
Self-heating
Response Time
Measurement Current
Sensor Length
Sensor Location
Insulation Resistance
Sheath Material
Dimension
External Leads
Handle Dimension
Handle Temperature Range
Optional Calibration
Rev. 201609V3
-200 °C to 670 °C
Nominal 100 Ω
0.00385 Ω/ Ω/°C
±0.04 °C at -200 °C
±0.035 °C at 0 °C
±0.05 °C at 200 °C
±0.09 °C at 420 °C
±0.15 °C at 660 °C
±0.04 °C at 0 °C after 100 hours at 420 °C
±0.01 °C
±0.007 °C after 10 times thermal cycles from minimum to
maximum temperatures
<=0.01 °C
50 mW/°C
5 seconds for 63% response to step change in water moving
at 3 feet per second
0.5 mA or 1 mA
32 mm
5 mm from tip
>1000 MΩ at room temperature
tm
Inconel
0.25 inch X 12 inch (6.35 mm X 305 mm)
tm
Teflon
–insulated copper wire, 4 leads, 2.5 meters
15 mm (OD) X 65 mm (L)
-50 °C to 180 °C
NIST traceable calibration and data available per request:
1660-12: Ordering # 5007
4/10