AccuMac AM1660 User Manual - Page 5
Browse online or download pdf User Manual for Thermometer AccuMac AM1660. AccuMac AM1660 10 pages. Precision industrial platinum resistance thermometer
AM1660 Precision Industrial PRT User's Guide
2 Specifications
2.1 Specifications
Temperature Range
Resistance at 0 °C
Temperature Coefficient
Accuracy
Drift
Short Term Stability
Thermal Shock
Hysteresis
Self-heating
Response Time
Measurement Current
Sensor Length
Sensor Location
Insulation Resistance
Sheath Material
Dimension
External Leads
Handle Dimension
Handle Temperature Range
Optional Calibration
Rev. 201609V3
-200 °C to 670 °C
Nominal 100 Ω
0.00385 Ω/ Ω/°C
±0.04 °C at -200 °C
±0.035 °C at 0 °C
±0.05 °C at 200 °C
±0.09 °C at 420 °C
±0.15 °C at 660 °C
±0.04 °C at 0 °C after 100 hours at 420 °C
±0.01 °C
±0.007 °C after 10 times thermal cycles from minimum to
maximum temperatures
<=0.01 °C
50 mW/°C
5 seconds for 63% response to step change in water moving
at 3 feet per second
0.5 mA or 1 mA
32 mm
5 mm from tip
>1000 MΩ at room temperature
tm
Inconel
0.25 inch X 12 inch (6.35 mm X 305 mm)
tm
Teflon
–insulated copper wire, 4 leads, 2.5 meters
15 mm (OD) X 65 mm (L)
-50 °C to 180 °C
NIST traceable calibration and data available per request:
1660-12: Ordering # 5007
4/10