Accurion halcyonics i4medium Panduan Petunjuk Pengoperasian - Halaman 5
Jelajahi secara online atau unduh pdf Panduan Petunjuk Pengoperasian untuk Peralatan Laboratorium Accurion halcyonics i4medium. Accurion halcyonics i4medium 13 halaman. Active vibration isolation desktop unit
5. Getting Started and Operation
Setting Up the System
To obtain best performance from the halcyonics_i4, set it up on a stable, rigid and flat surface. For
optimal operating results, the planarity of the support surface should be 0.5 mm.
If you set up the i4 system on a table or bench frame, make sure that the table is distortion-free and stiff.
Although most tables and bench frames have sufficient resistance to vertical vibration, they are relatively
sensitive with respect to horizontal vibration. This causes amplification of structure and airborne noise.
Therefore, to set up your vibration isolation equipment, select a place with a vibration level that is as low
as possible. Vibration generated at this place should not exceed velocities of 500 μm/s.
Operating the system at low temperatures may cause malfunctioning. If the equipment is brought
from a cold environment into a relatively warmer one, we recommend that you wait approx. 2 – 3
hours before plugging it into AC power and switching on the power.
Start-up
Set up the system on a flat, stable surface at your work area.
Connect the external power supply to AC power (main supply) and to the i4.
Center your application on the top plate of the i4 system.
Turn on the power switch on the front panel of the system.
Press the Load Adjust button for automatic load adjustment (see page 6)
Set the switch for active isolation to on, blue LED is on (see page 7)
Power switch
Isolation switch Load adjust switch
- 5 -