Accurion halcyonics i4medium Panduan Petunjuk Pengoperasian - Halaman 7

Jelajahi secara online atau unduh pdf Panduan Petunjuk Pengoperasian untuk Peralatan Laboratorium Accurion halcyonics i4medium. Accurion halcyonics i4medium 13 halaman. Active vibration isolation desktop unit

The system may only be transported or moved in the transport-locked condition!
There must not be any load on top of the isolation system during transport!
To change the different modes, it is necessary to press and hold the button for some seconds.

Using Active Vibration Isolation

Once you have started up the system, press the "isolation" button on the front panel to enable the active
vibration isolation. Now the system initializes, which is shown by a flashing blue LED. When the system
initializes for the first time or after a long interval, this process can take up to 30 seconds. When the blue
LED remains lit, the active vibration isolation is activated.

Overmodulation of the System

The i4 has been designed to compensate vibration amplitudes up to 500 µm/s. If vibrations significantly
exceed this level the system changes to the stand-by mode, indicated by a flashing blue "Isolation" LED.
After the overload excitation is stopped, the isolation mode will automatically be turned on again. After a
severe overload the system may take up to 30 seconds to reach full active isolation performance, but
normally only a few seconds are required.
The active vibration isolation will be automatically switched off during overmodulation of the
system. Once this interference has subsided, the system will re-initialize and, after a few seconds,
automatically resume to the active isolation mode. This procedure does not require any action
from the user.
- 7 -