Product Cleaning Instructions

In order to ensure accurate measurements, it is recommended
to clean the device after every use.
Please use a cotton swab to dip 75% alcohol to wipe the
thermometer probe and remove any residue.
We recommend you to clean the probe for at least for five
seconds.
Then wipe the other components with a soft and dry cloth.
After cleaning, please ensure that
there is no visible dirt or residue on
your thermometer. Wait 10 minutes
before using.
The
device
can
for maximum of 40000 times.

Troubleshooting

Problem
When the power is
turned on, the
screen doesn't
display.
There may be dirt on
the sensor.
09
be
cleaned
Ensure the battery
polarity is the same as
the battery case.
Clean the device.
The interval for each
measurement should
be over 10 seconds.
Reasons
When the temperature is
higher than the
1. When the measurement
measurement range (more
than 43.0°C/109.4°F for
the forehead mode, and
over 93.2°C/199.76°F in
the object mode), the
screen will display "HI"
2. When the measurement
When the temperature is
lower than the
measurement range (less
than 34.0°C/93.2°F for the
forehead mode, and below
0°C/32°F in the object
mode), the screen will
display "Lo".
The limit of the operating
environment is 40°C/104°F
When using this product, the
for this device. When the
environment temperature cannot
environment temperature
be higher than 40°C/104°F.
exceeds this temperature,
the screen will show the
error "Er.H".
The lower limit of the
When using this product, the
operating environment is
environment temperature cannot
10°C/50°F for this device.
be lower than 10°C/50°F.
When the environment
temperature is below this
temperature, the screen
will show the error "Er.L".
When the environment
temperature changes
When the error "Err" is
rapidly for 5 degrees in
displayed, place the device in a
object mode, the screen
stable environment, and do not
will show the error "Err",
use for 30 minutes.
and the device will
automatically shutdown.
Solutions
probe is not properly placed
on the forehead, or the
measuring distance is too
far, the measured result may
be low.
probe is dirty, the measured
result may be low. Please
clean the probe following the
previous cleaning
instructions.