ACD Cal 2000 Buku Petunjuk - Halaman 11

Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Alat Ukur ACD Cal 2000. ACD Cal 2000 13 halaman. Gas generator release vi

You may adjust the flow upwards by pressing the POWER
switch (arrow up) or downwards by pressing the SELECT
switch (arrow down). If you set the flow to 0.0 LPM, the
instrument will disable the pump and enter the flow draw
mode. Please refer to section III. Normal Operation:
Sample Draw Mode for more information. If you set the
flow between 0.2 and 1.0 LPM, the instrument will automati-
cally adjust the pump to the requested flow rate using
feedback from the built-in mass flow sensor.
Immediately after the flow rate screen appears, the ppm
screen appears.
Output: 000 ppm
SELECT to change
You may change the ppm by pressing the SELECT switch.
10 to 50 ppm
set . . . 000 ppm
You may adjust the ppm upwards by pressing the POWER
switch (arrow up) or downwards by pressing the SELECT
switch (arrow down). If you set the ppm to 000 ppm, the
instrument will not generate any gas, although it will
continue to pump.
The minimum and maximum ppm settings are dependent
upon the flow rate and the type of generating cell installed in
the instrument. The Cal 2000 will read the memory chip
inside the generating cell and automatically display the
minimum and maximum ppm available for the flow rate that
the instrument is set to. The total range of the cell is divided
into three sub ranges to allow for appropriate resolution in
each sub range. For all cells (excepting CLO
Micro-cells) the sub ranges at .5 LPM are .5 to 1 PPM,
and the
2
Page 9
IV.

Menu Options

Foreign Language Option
The menu options can be adjusted to read in German,
French, or Spanish. To change the language, start the unit as
you normally would. When the ACD screen appears,
A. C. D. VI
Copyright 2000
press the SELECT button. This will bring up the following
screen:
SELECT to change
Press the SELECT button to choose the preferred language.
When the language is displayed, press the POWER button to
continue with the start up sequence.
ClO2 Cell Reminder Screen
After the Battery Status screen appears during the start up of
the instrument, if a ClO2 cell is in the instrument, a reminder
message will appear to change the Chlorine Dioxide Electro-
lyte.
Electrolyte Exp.
Renew Often
Please read section VI. Cell Maintenance for detailed infor-
mation on changing your electrochemical generating cell's
electrolyte.
Page 14
English