ACD Cal 2000 Buku Petunjuk - Halaman 13

Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Alat Ukur ACD Cal 2000. ACD Cal 2000 13 halaman. Gas generator release vi

At any time you may reset either the flow rate or the ppm rate
by pressing the SELECT button and entering the status
screens as described above.
When you are finished generating gas and wish to turn off
the instrument, press and hold the POWER switch until the
purge screen is displayed.
]
Purging ...
Standby: 300 sec
The instrument will purge itself of gas for 300 seconds and
then automatically shut off. At any time during this period,
you may turn the instrument back on by pressing the
SELECT switch and entering the status screens as described
above. You must change either the flow rate or the ppm
setting and then the instrument will immediately go to the
stabilization screen and continue to run at the new settings.
The amount of stabilization time will depend upon how long
the instrument was in the purge mode.
Page 11
III. Normal Operation: Sample Draw Mode
The instrument may also be operated in a sample draw
mode where the internal pump is disabled and the air is
drawn through the instrument by an external pump. This
mode is used for calibrating sample draw instruments.
To enter the sample draw mode, press the SELECT switch
when the flow status screen is displayed.
Flow: 0.5 lpm
SELECT to change
When the flow set screen appears, adjust the flow to read
0.0 LPM.
0.2 to 1.0 lpm
set . . . 0.0 lpm
The ppm output screen will then read 000 ppm. You will be
unable to set the ppm until the sample draw instrument has
been attached to the outlet hose and a stable flow rate has
been established. While the instrument is waiting for the
sample draw unit to be attached, it will display the flow
stabilization screen.
Stabilize flow..
Flow: 0.000 lpm
Page 12