D400 Rear Panel Components
USB Socket
Potential
Equalisation
Connector (PEC)
Mains Inlet, Fuse Holder & Power Switch
Accepts standard IEC mains lead (supplied). Operates over 100-120V and 200-240V.
The D440 is fitted with fuses appropriate for the voltage selector setting. These fuses
should only be replaced by qualified personnel.
100-120 V mains T 400 mA L 5 x 20 mm, 250 V
200-240 V mains T 200 mA L 5 x 20 mm, 250 V
Voltage Selector
Check that the voltage selector on the back panel of the instrument is set correctly for your
supply voltage. The factory setting is 230 volts and the alternative is 115 V.
The 230 V setting allows use on supplies from 200 to 240 V.
The 115 V setting allows use on supplies from 100 to 120 V.
Potential Equalisation Connector (PEC)
Earth/Ground reference for unit and bonding point. This is to be used when the
earth/ground conductor in the mains lead cannot be relied upon.
USB Socket
Provides USB connection to host computer running Windows OS. USB connection allows:-
1.
Software control of D400 settings via Control Panel.
2.
Updates to firmware.
3.
Recalibration of the D400.
Voltage Selector
15
Version 1.2
Mains Inlet,
Fuse Holder &
Power Switch