AEMC instruments JM852 Panduan Pengguna - Halaman 3

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Alat Ukur AEMC instruments JM852. AEMC instruments JM852 4 halaman. Ac current probe

Overload: 3600A during 5 mn
Accuracy: Per IEC 185-26-27
Class 0.5 from 48 to 1000 Hz
Frequency Range:
30 Hz to 5 kHz; current derating above
1 kHz for continuous use
Ampere Second Product: 90 AA
Load Impedance:
Working Voltage: 600V AC
Common Mode Voltage: 600V AC
Influence of Adjacent Conductor:
0.005A/A AC
Influence of Conductor in Jaw Opening:
1% ± 0.1A of reading
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 122°F (-10° to 50°C)
Storage Temperature:
-13° to 176°F (-25° to 80°C)
Influence of Temperature:
< 0.2% per 10°K
Jaw Opening: 3.54" (90mm)
Maximum Conductor Size:
Cable: 2.52"  max. (64mm)
Bus bar: 1.97 x 5.31" (50 x 135mm)
2.52 x 3.94" (64 x 100mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
500mm (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15 mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Jaws: ABS UL 94 V2
100K
Dimensions:
4.72 x 12.40 x 1.89" (120 x 315 x 48 mm)
Weight:
2.65 lbs. (1200 g)
Colors:
Dark gray handles with red jaws
Output:
5 ft (1.5m) lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between primary or secondary and outer case of
handle upon IEC 1010-2-32
Common Mode Voltage:
600V CAT III, Pollution: 2
ORDERING INFORMATION
Current Probe JM852 ................... Cat #2113.41
Accessories:
Banana plug adapter
(to non-recessed plug).................... Cat #1017.45