Alcolizer LE 5 Buku Petunjuk - Halaman 8
Jelajahi secara online atau unduh pdf Buku Petunjuk untuk Peralatan Uji Alcolizer LE 5. Alcolizer LE 5 11 halaman.
Figure 5 - Self check underway (Please Wait...) and Ready messages
Press the Test key to start the calibration sequence. The device displays a
countdown sequence: 13 - 12 - 11 - 10 etc. at one second intervals.
Turn on the calibration gas flow when the count is approximately 10 or 9.
When the countdown reaches 0, the device will take a sample of the calibration gas
and a beep sound will be heard. The Analysing Sample screen will be displayed
(refer Figure 6).
Turn off the calibration gas flow after the device has taken a sample and the beep
tone has ceased.
If the calibration process is unsuccessful, error code 43 will be displayed. Consult Alcolizer
Technology for further technical advice on 1300 789 908.
REF: LE5 Calibration Instructions_Whole Units V5
Figure 6 - Analysing Sample message
NOTE
Page 8
© Alcolizer Technology