9 Displaying Calibration Results
The results of the calibration are displayed on the screen. The table below describes what
each message means.
The new GAIN needs to be recorded on the data sheet.
Screen Message
Pre-Cal Result
Displays the calibration gas reading for the
old gain figure value
New Gain
Displays the new calibration gain figure
value
Reference Gas
The calibration reference gas value (as
entered above)
Backup Battery
Displays the backup battery voltage level.
Recovering
Displays the fuel cell recovery period
Press Key to
Press either the Test or Up/Down keys to
continue
return the device to the calibration menu
display screen
If the device has been used or calibrated within the last 24 hours, a calibration lockout message will display,
preventing calibration of device for the period displayed (refer
Turn device off and then on.
Enter the 'About' menu and press the main button until the unit's details are
displayed.
Check if device displays the correct remain days to next calibration.
REF: LE5 Calibration Instructions_Whole Units V5
Description of Results
NOTE
Figure 7 - Calibration Lockout message).
Figure 7 - Calibration Lockout message