Direct Supply Panacea 3250 Panduan Pemeliharaan - Halaman 4

Jelajahi secara online atau unduh pdf Panduan Pemeliharaan untuk Peralatan Medis Direct Supply Panacea 3250. Direct Supply Panacea 3250 8 halaman. Soft touch siderails for beds
Juga untuk Direct Supply Panacea 3250: Panduan Pemilik (12 halaman)

Direct Supply Panacea 3250 Panduan Pemeliharaan

15-Point Bed Inspection

Step 1:
Perform the procedure "STEPS
TO PREVENT BED FIRES" listed in the
bed manual.
Step 2:
Review and inspect for compliance to
the "WARNINGS" listed in the bed manual.
Step 3:
Review "REDUCING THE RISK OF
ENTRAPMENT" listed in the bed manual.
Step 4:
Check casters to ensure they lock, if
applicable, and roll properly.
Step 5:
Inspect all bed components for
damage or excessive wear.
4
Step 6:
Visually examine all welds for cracks.
Step 7:
Inspect the deck components for
bending or damage.
Step 8:
Check the motor actuator shaft
and its connections for bending, damage or
excessive wear.
Step 9:
Check actuator ends and its mounting
hardware for bending or excessive wear .