- ページ 4
医療機器 Direct Supply Panacea 3250のPDF メンテナンス・マニュアルをオンラインで閲覧またはダウンロードできます。Direct Supply Panacea 3250 8 ページ。 Soft touch siderails for beds
Direct Supply Panacea 3250 にも: オーナーズマニュアル (12 ページ)
15-Point Bed Inspection
Step 1:
Perform the procedure "STEPS
TO PREVENT BED FIRES" listed in the
bed manual.
Step 2:
Review and inspect for compliance to
the "WARNINGS" listed in the bed manual.
Step 3:
Review "REDUCING THE RISK OF
ENTRAPMENT" listed in the bed manual.
Step 4:
Check casters to ensure they lock, if
applicable, and roll properly.
Step 5:
Inspect all bed components for
damage or excessive wear.
4
Step 6:
Visually examine all welds for cracks.
Step 7:
Inspect the deck components for
bending or damage.
Step 8:
Check the motor actuator shaft
and its connections for bending, damage or
excessive wear.
Step 9:
Check actuator ends and its mounting
hardware for bending or excessive wear .