ΩHMRANGER-LCV - BASIC MENU FUNCTIONS
Press the Menu button to exit measuring and
enter the Menu mode. This screen will
appear. Use the Navigate and Select buttons
to move through the Menu. The options are:
•
Measure
•
Voltage
•
Powerdown
•
LCD Light
•
4 Beep Read
•
Details
VOLTAGE - Use this Menu option to select the
maximum stimulus voltage that will be applied
by ΩhmRanger-LCV.
The options are available are:
•
•
POWERDOWN - Use this function to set and configure the period of
inactivity after which ΩhmRanger-LCV will power down automatically.
The options are:
•
•
•
•
•
•
DETAILS - Will display the ΩhmRanger-LCV's unique serial number,
calibration date and status, and software version.
USING THE FULLY ISOLATED USB INTERFACE
ΩhmRanger-LCV's micro-USB interface can be used to export
readings and to control the device via a PC (SCPI-format). The
USB interface
cannot
be used to power ΩhmRanger-LCV, as it is
isolated from the main circuitry to prevent interference. For help
connecting ΩhmRanger-LCV, see the guidance in our website's
ΩhmRanger-LCV section. For SCPI instructions, please download
the full OR-01 Instruction Manual at:
USE THIS QUICK START GUIDE IN CONJUNCTION WITH THE FULL INSTRUCTION MANUAL. DOWNLOAD AT:
380 mV (the default setting), or
206 mV
Off
30 Minutes
1 Hour (the default)
2 Hours
4 Hours
8 Hours
LCD LIGHT - Use this function to toggle
the ΩhmRanger-LCV screen backlight
ON and OFF. Disabling the backlight will
extend battery life.
4 BEEP READ - Use this function to
take readings only when Resistance
values are stable within selected
tolerances. An audible prompt is given
when this is achieved.
www.altonovus.com
MODEL: OR-01: LOW COMPLIANCE VOLTAGE OHMMETER.
QUICK START GUIDE (Rev 1.0 dated 10 MAR 22)
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the User is encouraged to try to correct the interference by one or more of
the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different that to
which the receiver is connected.
- Consult the dealer, or an experienced radio/TV technician for help.
WWW.ALTONOVUS.COM