CUSTOM ULTRASONICS, INC.
Label
Technical Specifications
Weight: 175 lbs.
Overall height: 93"
Overall width: 36"
Depth: 9¾"
Electrical Rating: 120 VAC, 5A, 60 Hz, 1/8 HP
Hospital grade plug: The Storage Cabinet is supplied with 15A, 125 VAC, NEMA 5-15 Type
Power cord: 13/3, 600V, 90°C, SJTO
Environmental Operating Conditions
a. Intended for indoor use.
b. Temperatures 41°F to 104°F (5°C to 40°C).
c. Relative Humidity of 20% to 80%.
d. Mains supply voltage fluctuations up to ±10% of the nominal voltage.
Physical Description
The Endoscope Storage Cabinet, Model 20000, is fabricated of powder coated steel to provide an
environment for endoscope storage that is free of water-absorbent materials.
A gasket surrounds the door opening. When the door is closed (by the adjustable and lockable handle),
the gasket is compressed providing a secured enclosure.
At the top of the Storage Cabinet are two compartments.
o One is an enclosure containing an air pump for the circulation of air through the channels of the
endoscopes.
o Next to the air pump, on the other side of the upper enclosed area, is a chamber for desiccant bags
that are used to decrease the level of moisture from the re-circulated air.
Two re-sealable plastic containers, each containing 6 desiccant bags, are provided with the
initial Storage Cabinet purchase.
Two packages of (3) Humidity Indicator Cards are also provided.
MAN-009
Graphical Symbol: Indicates protective earth ground.
Graphical Symbol: Indicates an earth ground.
Graphical Label: Indicates the device AC power connection rating.
Graphical Label: Indicates the device has been tested and conforms
to UL STD 61010-1 Certified to CAN/CSA STD C22.2 NO.61010-1.
UDI Device Label: Indicates the device and production information
that corresponds to the specific device through plain-text and barcode
formats. This label is applied directly to the device.
Revision 08232021
Endoscope Storage Cabinet
Model 20000
Operator's Manual
Indication
Page 4 of 11