CUSTOM ULTRASONICS, INC.
A series of perforated holes, located on the bottom of the desiccant chamber, allows for air movement
through the desiccant and into the air pump for re-circulation of air through the endoscopes' internal
channels.
By opening the desiccant chamber door, the six (6) bags of desiccant and (3) humidity indicator cards
can be monitored and replaced with six (6) new bags of desiccant and (3) new humidity indicator cards.
Below the desiccant bags and air pump chambers, at the top of and inside the main chamber of the
cabinet, is a manifold containing four (4) bulkhead fittings to accommodate eight (8) endoscopes total.
o Adapters are used to connect the endoscope(s) to these four (4) dual bulkhead fittings. The
adapters are provided by Custom Ultrasonics and require using OEM parts from endoscope
manufacturer when applicable.
o Circulation of air through the endoscopes' internal channels occurs through these adapters.
Although a relief valve protects the air pump in the absence of adapters plugged into the bulkhead
fittings.
CAUTION: One adapter must remain in the adapter manifold at all times to prevent damage to
the air pump.
Accessory trays for the endoscopes' operating buttons are located inside the door of the Storage
Cabinet.
The Storage Cabinet's On/Off switch and the Adjustable Programmable Timer Dial are located outside
of the Storage Cabinet on the right front wall.
Each Endoscope Storage Cabinet, Model 20000, is equipped with a lock. Routinely locking the Storage
Cabinet may reduce the incidence of endoscope exposure and theft.
Note:
For proper operation of this Storage Cabinet, read and understand all instructions before using
the device. If the Storage Cabinet is used in a manner not specified by Custom Ultrasonics, the
protection provided by the device may be impaired.
MAN-009
Endoscope Storage Cabinet
Revision 08232021
Model 20000
Operator's Manual
Page 5 of 11