1-CUBE SU 6.2. Panduan Pengguna - Halaman 4
Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Termostat 1-CUBE SU 6.2.. 1-CUBE SU 6.2. 13 halaman. Water thermostat
4. T
ECHNICAL DATA
Electric data:
- voltage system TN-S 1+PE+N
- voltage 230V/50 Hz or 110V/60 Hz
- protection IP 20
- the equipment can be used in neutral medium
- the equipment output : hot bath 1500W, cold bath 1800W
- circuit breaker–10A fusible cut out
- plugging in el. network: with the help of flexible cord LYS 3x1,5
-averaged plug and socket outlet protected with 16A circuit breaker
Adjustable data:
- type of method: THERMOSTAT, PROFILE, CALIBRATION
- temperature range: from 25
and display
- time since the begining of test
Accuracy:
- accuracy of temperature measurement: +-0,01
- accuracy of temperature regulation: : + - 0,1
Data represented on LCD display:
- type of method: THERMOSTAT, PROFILE
- instantaneous temperature of bath or suspension, required temperature of bath
- time since the begining of test
- communication of the instrument with the service
0
0
C to +90
C, cold bath from - 10
0
C
0
C
0
0
C to +30
C with the help of membrane keyboard
4