AkraLab PLUGGO Buku Pegangan Pengguna - Halaman 2

Jelajahi secara online atau unduh pdf Buku Pegangan Pengguna untuk Peralatan Laboratorium AkraLab PLUGGO. AkraLab PLUGGO 15 halaman. Automated de-capper

1. General Information – Read this first
Type of equipment and intended use:
1.
Pluggo is used for safe removal of original caps from vacuum collection tubes. It is
intended for use in laboratory environment under normal in-house conditions.
Clinicon AB cannot accept any responsibility for any loss or damage, however caused,
arising from a faulty or incorrect use of the instrument.
Tube Handling
2.
Pluggo is capable of handling 5ml, 7ml, 7ml long and 10ml Vacuum tubes of the same
brand.
Note: Each tube size has to be loaded in its corresponding carousel size with different
colour marks.
Note: Never load de-capped tubes in carousel. Pluggo will damage tubes without caps!
Note: Never change brand without consulting Pluggo technical support.
Carousels
3.
Different tube sizes are handled by use of different carousels with corresponding colour
marks.
Carousel colour
Blue
Red
Grey
Yellow
Note: All tubes placed in a carousel must be of the same type and size. Otherwise,
damaged tubes may result.
Each carousel is capable of loading up to 24 tubes.
[email protected] · 965 116 521 · 902 222 275 · www.akralab.es
Tube size
13 x 75 mm
13 x 100 mm
16 x 75 mm
16 x 100 mm
Volume
5 ml
7 ml long
7 ml short
10 ml
2