Ashcroft DG25 Panduan Petunjuk Pemasangan dan Pengoperasian - Halaman 10

Jelajahi secara online atau unduh pdf Panduan Petunjuk Pemasangan dan Pengoperasian untuk Alat Ukur Ashcroft DG25. Ashcroft DG25 16 halaman. Digital pressure gauge

Ashcroft DG25 Panduan Petunjuk Pemasangan dan Pengoperasian
Light:
Determines how long the back light will remain
ON after any key is pressed in Measurement or
Menu Modes (Note: The timer is reset with any
key being pressed.)
1. While in Menu mode, press
until "LIGHT" is displayed.
©
k
2. Press
for timer value to appear;
"ON" refers to the back light remaining
ON at all times unit is powered ON,
"PRESS" designates that the back light
is switched on / off by briefly pressing
©
the backlight
M ENU
dicates the back light will never be illu-
minated. Selecting time values, "1 MIN"
(default), "5 MIN", "20 MIN", will acti-
vate the backlight symbol on the LCD
display for the designated minutes.
©
k
3. Press
to select back light time
(display will briefly show "DONE") and
return to Measurement Mode.
Note: In "PRESS" mode; factory has set 1
hour timer to save battery.
Feature: Back light indicator will blink 10 sec-
onds prior to light shut off.
©
©
,
whereas, "OFF" in-
10
Update:
Utilized to select the rate at which the displayed
pressure value is updated on the screen. This
function is used when rapid changes in pres-
sure cause "flutter" in the display values;
longer intervals will reduce the update rate and
"average" the readings on such applications.
1. In Menu mode, press
"UPDAT" appears.
©
k
2. Press
to select.
©
©
3. Press
,
"1 SEC" (default), "500 MSEC", or "250
MSEC".
©
k
4. Press
to select value (display will
briefly show "DONE") and return to
Measurement Mode.
Note: Changing value to anything other than
1 sec may cause a slight zero offset, and it
is recommended that the gauge be fully
vented and re-zeroed before taking accurate
readings. Also, battery life will be reduced
by use of an update rate faster than 1 SEC.
©
©
,
until
to select values for