Akron 8601B Panduan Pengoperasian - Halaman 11

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Peralatan Medis Akron 8601B. Akron 8601B 20 halaman. Childs tilttable
Juga untuk Akron 8601B: Panduan Servis (14 halaman)

4. Cleaning and Decontamination

Do not use Alkalis or other organic solvents for routine cleaning.
Always cover vinyl with fresh paper or sheet between patients.
If contamination is known or suspected always decontaminate couch before use.
Frequency of routine couch cleaning will depend on exact application.
CLEANING / DECONTAMINATION
Wear suitable protective clothing
Disconnect mains supply
Locate couch in a suitable cleaning area with castors disengaged or wheels locked.
A)
Using a clean damp cloth, warm water and soap carefully wipe couch from top to bottom, taking
care around electrical components.
B)
Wipe couch carefully from top to bottom using a damp cloth and clean water.
C)
Wipe couch carefully dry with clean paper towels from top to bottom.
If DECONTAMINATING then repeat steps a) to c) using a dilute Hypochlorite / warm water mixture
(10,000ppm) instead of warm water and soap.
Safely discard cleaning / decontaminating materials.
Wash hands thoroughly.
Put couch back into service.
WARNING
Replace torn or damaged vinyl immediately.
AKRON 8601B
9