4. Cleaning and Decontamination
Decontamination of Straps
It is not recommended that the straps are laundered on a regular or routine basis as this will eventually
compromise the integrity of the material and associated Tac-Flex. However, under certain conditions, ie
when the straps become heavily soiled or when they are suspected or known to have been in contact with
an infectious organism, they can be laundered or decontaminated using one of the following methods.
General Wash
The straps should be washed separately, as the Tac-Flex will attract debris from other material which will
ultimately reduce its effectiveness.
The Tac-Flex should be joined prior to washing.
The straps may then be laundered in the normal fashion, at a temperature of approximately 71°C and
allowed to dry in ambient conditions.
After such wash, the straps should be thoroughly inspected prior to patient use. If deterioration is present
or the Tac-Flex fails to maintain integrity, the straps must be replaced.
Decontamination
Eye protection and suitable clothing should be worn. Where contamination is known or suspected, the
surface of the straps should be sprayed liberally using a hypochlorite solution of 1,000 ppm or a 1,000
ppm solution of sodium dichloroisocyanurate (NaDCC), after removing excessive organic material. The
straps should then be air dried at ambient temperature and inspected (as above) prior to patient use.
5. Troubleshooting – Electric Actuator
Symptom
Actuator will not work
If the problem persists consult your local Akron Supplier.
AKRON 8601B
Possible cause
Power disconnected
Fuse blown
Hand/footswitch
disconnected
Ensure mains supply is connected
Replace fuse in socket
Check that the hand/foot switch is fully
inserted into the actuator
10
Action