Dolomite Mitos Manual Petunjuk Pemrograman - Halaman 11

Jelajahi secara online atau unduh pdf Manual Petunjuk Pemrograman untuk Pompa Air Dolomite Mitos. Dolomite Mitos 15 halaman. Automated product collector
Juga untuk Dolomite Mitos: Panduan Memulai Cepat (3 halaman)

Dolomite Mitos Manual Petunjuk Pemrograman
The Dolomite Centre Ltd.
Field
Qc
Current flow rate in pl/s. Requires a connected Mitos Sensor Display flow meter. Note
the actual accuracy is given by flow sensor type below. Contact Dolomite Support for
more information on this feature.
Qt
Target flow rate in pl/s. Requires a connected Mitos Sensor Display flow meter. Contact
Dolomite Support for more information on this feature.
Ft
Flow sensor type plus flow control mode. Requires a connected Mitos Sensor Display or
sensor interface flow meter. This consists of an upper nibble, middle nibble and lower
nibble:
• The high nibble describes flow control mode: 0x100 means it is flow control or
0x000 otherwise. This is only meaningful when the pump is controlling mode.
• The middle nibble describes the connection to the Flow Sensor: 0x10 means it is
connected to the Sensor Display Module (3200095); 0x00 means it is connected to
the Sensor Interface Module (3200200)
• The lower nibble is defined as (Flow Sensor Types, see 7.1).

6.4 Error Codes

If the pump state is ERROR, you can get the exact error code, date and time and limited
description by using the "e" command. The table below shows the possible values of the error
codes that may be returned from the pump when the pump state becomes ERROR.
Code
Meaning
0
None
1
Supply larger
than maximum
2
Tare time out
3
Tare supply still
connected
4
Control start
time out
5
Pressure target
too low
6
Pressure target
too high
7
Leak test supply
pressure too
low
8
Leak test time
out
100
Broken
3600051 - P-Pump API - programming instructions - v 1.3
Description
There is no current error.
The connected supply pressure is too high for the device (>11.5bar).
The tare did not complete successfully, no steady base line detected
The microcontroller cannot tare because the supply pressure is still
connected.
This error occurs if open/closing the valves does not change the
chamber pressure within a given time period, suggesting something is
mechanically wrong (valve stuck/broken?). Contact Dolomite Support.
The pressure target set or required is too low, below the current
minimum pressure for the current supply pressure.
The pressure target set or required is too high, above the current
maximum pressure for the current supply pressure.
The pressure is too low (<400mb) for leak test.
Leak test times out if either the pressure cannot reach target or after
reaching target is not stable in 30 seconds
The main board has failed to communicate with the microcontroller
board. This indicates something is very wrong internally with the
pump. Contact Dolomite Support.
Description
Page 11 of 15