Dolomite Mitos Manual Petunjuk Pemrograman - Halaman 13

Jelajahi secara online atau unduh pdf Manual Petunjuk Pemrograman untuk Pompa Air Dolomite Mitos. Dolomite Mitos 15 halaman. Automated product collector
Juga untuk Dolomite Mitos: Panduan Memulai Cepat (3 halaman)

Dolomite Mitos Manual Petunjuk Pemrograman
The Dolomite Centre Ltd.
seen above, so when converting these values for display you need to convert the units from pl/s
to the units for the flow sensor in the table above and then pay attention to the number of
decimal places it supports.
7.2 Fluid names
Note the minimum flow rate actually means the minimum calibrated flow rate, lower values can
measured but will be less accurate.

8 Programming Example

The table below shows a brief example of a set of commands and the responses produced using
pressure control. The example starts the pump, tares it and then controls it at pressure before
generating and clearing an error. The termination characters are omitted for clarity.
Cmd.
Response
s
#s0,0,0,-2,-3,0,0,0,0
A1
#A0
s
#s0,0,1,-2,-3,0,0,0,0
R1
#R0
s
#s0,2,1,-2,-3,0,0,0,0
s
#s0,0,1,0,0,0,0,0,0
s
#s0,0,1,0,7500,0,0,0,0
P2000
#P0
s
#s0,1,1,1589,7500,2000,0,0,0
s
#s0,1,1,2001,7500,2000,0,0,0
P8000
#P0
s
#s6,3,1,27,7562,8000,0,0,0
e
#eMon Aug 6 10:38:18 2012:Error on ppbLoglet: 6, Target beyond range
C
#C0
s
#s0,0,1,2,7500,0,0,0
A0
#A0
s
#s0,0,0,7500,0,0,0
3600051 - P-Pump API - programming instructions - v 1.3
Fluid
Short Name
Water
<SPACE>H2O
FC-40
FC40
Novec 7500
NOVE
Hexadecane
HEXA
Mineral Oil
<SPACE>OIL
OK – manual mode, no connected supply
Entering remote mode, command accepted
OK – remote mode and IDLE
Start TARE, command accepted
Taring....
... wait a few seconds for tare to...
...succeed. Back to IDLE. Now connect supply
OK
Set target to 2000 and CONTROL, command
accepted
...pressurising...
... and controlling at 2000mbar
Attempt to set target too high...
... results in ERROR state, so use "e" to read the error
then clear it
Error state cleared and return to IDLE
OK
Back to manual...
... in IDLE and vented
Comment