Baumer CleverLevel LBFS Series Manuale di istruzioni per l'uso - Pagina 13
Sfoglia online o scarica il pdf Manuale di istruzioni per l'uso per Accessories Baumer CleverLevel LBFS Series. Baumer CleverLevel LBFS Series 48. Level measurement, point level detection
11. Cleaning, maintenance and repair
Cleaning
Clean, disinfect or sterilize sensor as needed
f
(CIP/SIP).
Maintenance
Regular maintenance is not required.
12. Disposal
Do not dispose of in household waste.
f
Separate materials and dispose of in
f
compliance with nationally applicable
regulations.
13. Accessories
For adapter and other accessories see www.baumer.com.
14. Technical data
Environmental conditions
Operating temperature range
Storage temperature range
Ambient humidity
Protection class
Oscillations (sinusoidal)
(EN 60068-2-6)
Power supply
Voltage supply range
Reverse polarity protection
Current consumption
(without load)
Power-up time
Features
Repeatability
Hysteresis
Response time
Damping
www.baumer.com
–40...+85 °C
–40...+85 °C
< 98 % RH,
condensing
IP67
IP69K (with
appropriate cable)
1.6 mm p-p
(2...25 Hz),
4 g (25...100 Hz),
1 octave / min.
12...30 V DC
yes
25 mA typ.,
50 mA max.
< 2 s
± 1 mm
± 1 mm
0.1 s
0.0...10.0 s
(configurable)
Operating Instructions
Level measurement
CleverLevel
Point level detection
Repair
Do not repair the sensor yourself.
Send damaged sensor to Baumer.
f
Output signal
Output type
Current load
Short circuit
protection
Voltage drop
Leakage current
Switching logic
Cable version
Operating tempera-
ture range
Bending radius min.
®
LBFS
PNP
NPN
20 mA max.
yes
PNP: (+Vs –1.5 V) ± 0.5 V,
Rload = 10 kΩ
NPN: (+1.5 V) ± 0.5 V,
Rload = 10 kΩ
± 100 μA max.
Normally open (NO),
active low
Normally closed (NC),
high enabled
–25...+70 °C (if cable is not
moved)
–5...+70 °C (if cable is
moved)
r ≥ 10 mm
13 / 48