Baumer CleverLevel LBFS Series Operating Instructions Manual - Page 13

Browse online or download pdf Operating Instructions Manual for Accessories Baumer CleverLevel LBFS Series. Baumer CleverLevel LBFS Series 48 pages. Level measurement, point level detection

Baumer CleverLevel LBFS Series Operating Instructions Manual

11. Cleaning, maintenance and repair

Cleaning
Clean, disinfect or sterilize sensor as needed
f
„
(CIP/SIP).
Maintenance
Regular maintenance is not required.

12. Disposal

Do not dispose of in household waste.
f
„
Separate materials and dispose of in
f
„
compliance with nationally applicable
regulations.

13. Accessories

For adapter and other accessories see www.baumer.com.

14. Technical data

Environmental conditions
Operating temperature range
Storage temperature range
Ambient humidity
Protection class
Oscillations (sinusoidal)
(EN 60068-2-6)
Power supply
Voltage supply range
Reverse polarity protection
Current consumption
(without load)
Power-up time
Features
Repeatability
Hysteresis
Response time
Damping
www.baumer.com
–40...+85 °C
„
„
–40...+85 °C
„
„
< 98 % RH,
„
„
condensing
IP67
„
„
IP69K (with
„
„
appropriate cable)
1.6 mm p-p
„
„
(2...25 Hz),
4 g (25...100 Hz),
1 octave / min.
12...30 V DC
„
„
yes
„
„
25 mA typ.,
„
„
50 mA max.
< 2 s
„
„
± 1 mm
„
„
± 1 mm
„
„
0.1 s
„
„
0.0...10.0 s
„
„
(configurable)
Operating Instructions
Level measurement
CleverLevel
Point level detection
Repair
Do not repair the sensor yourself.
Send damaged sensor to Baumer.
f
„
Output signal
Output type
Current load
Short circuit
protection
Voltage drop
Leakage current
Switching logic
Cable version
Operating tempera-
ture range
Bending radius min.
®
LBFS
PNP
„
„
NPN
„
„
20 mA max.
„
„
yes
„
„
PNP: (+Vs –1.5 V) ± 0.5 V,
„
„
Rload = 10 kΩ
NPN: (+1.5 V) ± 0.5 V,
„
„
Rload = 10 kΩ
± 100 μA max.
„
„
Normally open (NO),
„
„
active low
Normally closed (NC),
„
„
high enabled
–25...+70 °C (if cable is not
„
„
moved)
–5...+70 °C (if cable is
„
„
moved)
r ≥ 10 mm
„
„
13 / 48