Ankom TDF Series Procedura di assistenza - Pagina 8
Sfoglia online o scarica il pdf Procedura di assistenza per Strumenti di misura Ankom TDF Series. Ankom TDF Series 9. Vacuum sensor installation
Anche per Ankom TDF Series: Procedura di assistenza (7 pagine), Procedura di assistenza (4 pagine)
17.
Adjust the reported temperatures on the Touch Screen
Display with a calibrated thermometer accurate to at least
0.2˚C.
a. Obtain a calibrated thermometer and place it in the
immediate environment of the paddles.
b. Allow the temperature reading to equilibrate with the
environment.
c. Turn the TDF instrument on.
c-1. From the Select a Function screen, press Diagnostics
c-2. Service Mode
c-3. Temperature Sensor Corrections
c-4. Enter temperature offsets into the fields for each
station. For example, if the calibrated thermometer is
showing a temperature of 23.2˚C but the Uncorrected
temperature value for station 1 is showing a
temperature of 24.0˚C, enter an offset of -0.8˚C. If the
Uncorrected temperature value for station 2 is
showing a temperature of 23.0˚C, enter an offset of
0.2˚C. Whatever number is entered as the offset
(whether positive or negative) will be added to the
uncorrected value.
c-5. Write down the offset numbers you put in for each
Paddle Bar to later ensure they save to the instrument.
c-6. Select Save to Permanent Memory and exit out of
Diagnostics.
d. Turn the instrument off and then on again to check to
ensure the corrected value information saved properly.
e. Follow the above screen display selections again and check
to ensure the Temperature Sensor Corrections match those
that you wrote down after adjusting.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091
Paddle Bar Replacement
TDF
Service Procedure
Thermometer
TS021
Revised: 01/24/2017
pg. 8 of 9