16.
Secure the Shoulder Screw in place.
a. Tilt the instrument forward to access the back once
again.
b. Remove the Shoulder Screw installed finger-tight in the
previous step.
c. Apply a drop of Loctite 242 (provided in the new
Paddle Bar kit) to the threads of this screw.
d. Re-install the screw (do Not to get the Loctite on the
brass Rod End of the Mixer Piston) and tighten with the
⅛" Hex key wrench.
e. Return the instrument to its upright position.
f. Reinstall the two Service Access screws.
ANKOM Technology  2052 O'Neil Road  Email: [email protected] Phone: 315-986-8090  Fax: 315-986-8091
Paddle Bar Replacement
TDF
Service Procedure
Apply
Loctite
1/8"
Hex key
Wrench
Service
Access
Screws
TS021
Revised: 01/24/2017
pg. 7 of 9