- ページ 7

計測機器 Ankom TDF SeriesのPDF サービス手順をオンラインで閲覧またはダウンロードできます。Ankom TDF Series 9 ページ。 Vacuum sensor installation
Ankom TDF Series にも: サービス手順 (7 ページ), サービス手順 (4 ページ)

16.
Secure the Shoulder Screw in place.
a. Tilt the instrument forward to access the back once
again.
b. Remove the Shoulder Screw installed finger-tight in the
previous step.
c. Apply a drop of Loctite 242 (provided in the new
Paddle Bar kit) to the threads of this screw.
d. Re-install the screw (do Not to get the Loctite on the
brass Rod End of the Mixer Piston) and tighten with the
⅛" Hex key wrench.
e. Return the instrument to its upright position.
f. Reinstall the two Service Access screws.
ANKOM Technology  2052 O'Neil Road  Email: [email protected] Phone: 315-986-8090  Fax: 315-986-8091
Paddle Bar Replacement
TDF
Service Procedure
Apply
Loctite
1/8"
Hex key
Wrench
Service
Access
Screws
TS021
Revised: 01/24/2017
pg. 7 of 9