14.
Check the back stroke of the paddles.
a.
Return the instrument to the upright position.
b.
Using the depth gauge (provided in the new Paddle Bar
kit) with the Paddle Bar Assembly in the back stroke
position (this is its normal, at rest, position) rest the
squared-off end of the depth gauge against the bottom
edge of the #5 paddle.
c.
Measure the distance from the bottom edge of this
paddle to the face of the orange gasket on Clamp Bar B.
This distance should be 10/32". If this distance is not
correct contact ANKOM Technology for instructions on
how to make an adjustment.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091
Paddle Bar Replacement
TDF
Depth
Gauge
10/32"
Distance
4/32" – 5/32"
Distance
TS021
Service Procedure
Revised: 01/24/2017
10/32"
Distance
pg. 5 of 9