11.
Place the four shims in their appropriate positions.
Set the four 0.015" Shim spacers (provided in the new
a.
Paddle Bar kit) underneath the #1, 2, 5 & 6 paddles and
set on top of Clamp Bar B to properly distance the
Paddle Bar Assembly from Clamp Bar B. When doing
this, allow the paddles to rest on the shims while tilting
the paddles back so that the bottom front corner of each
paddle is the only part of the paddle that is touching the
shims. In this position the face of the paddles will not
be perfectly vertical but slightly tilted back at the
bottom.
b.
With the paddles held in a tilted back position, tighten
the four screws in the Paddle Bar Hinge Posts using the
5/32" Hex key wrench.
c.
Once the screws are tight, remove the four Shim spacers
from beneath the paddles.
12.
Plug the 25 pin Cable Connector into its socket in the
underside of the Electrical Enclosure.
a.
Carefully tilt the instrument forward.
b.
Make sure the cable is not caught up in the tubing or air
lines in the back of the instrument.
c.
When plugging the connector in, align the pins to the
sockets within the connector and twist clock-wise until
it stops turning.
d.
Check to make sure it cannot be pulled out (it should
lock in place).
13.
Finger-tighten the Shoulder Screw.
This will connect the cable block (on the Paddle Bar
Assembly sticking through the hole in the front of the
instrument) to the brass Rod End of the Mixer Piston in
order to secure the Paddle Bar while testing it in the
following steps.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091
Paddle Bar Replacement
TDF
Service Procedure
Shim
Spacers
underneath
Paddle
Align
Cable
Connector
Sockets
Shoulder
Screw
TS021
Revised: 01/24/2017
pg. 4 of 9