16.
Secure the Shoulder Screw in place.
a. Tilt the instrument forward to access the back once
again.
b. Remove the Shoulder Screw installed finger-tight in the
previous step.
c. Apply a drop of Loctite 242 (provided in the new
Paddle Bar kit) to the threads of this screw.
d. Re-install the screw (do Not to get the Loctite on the
brass Rod End of the Mixer Piston) and tighten with the
⅛" Hex key wrench.
e. Return the instrument to its upright position.
f. Reinstall the two Service Access screws.