After the unit has been set up and the user has become familiar with the System's control panel functions: 
1.
Attach the appropriate "E‐Type" handpiece to the motor. It can be used with most surgical contra‐angles without internal 
light or spray on the market. Use only handpiece that conform to ISO 3964. Change the handpiece and angle piece only when 
the motor not running. 
 
 
 
 
 
 
 
 
 
1.
Insert bag hanger rod into socket on the top of the console. 
2.
Attach the motor to the connector on the right side (facing the unit) on the front panel. 
3.
Attach the foot control to the connector (gray color) on the front panel. 
4.
Install irrigation tubing into pump on the rear panel. Use only tubing as described in "Irrigation Tube Set Up".  
5.
Switch on (on the rear panel. 
 
In every program Programs [ 1 to 5] are set the same parameters by the manufacturer. 
Perform  torque  calibration,  press  "Cal"  key  on  front  panel  while  holding  the  motor.  This  function  checks  the  handpiece  mechanical 
inertia, to off‐set this value. 
This operation will take a few seconds, the drill will run at different speed. 
The drill must not be engaged in any operation when this function is performed. 
Calibration should be done before each surgical session or anytime the handpiece is changed or lubricated.
I TECH LITE will store the operative parameters for each "program" as soon as the user sets them.  They will be maintained in the memory 
even if the central unit is switched off. For each "program" it is possible to set speed, torque, ratio, direction of rotation and flow. To 
reset the parameters of the manufacturer (SystemReset), press "RESET" key on Display screen N° 2. Wait until the end of the operation. 
 
9.1Programming the unit 
:
R/R
 Sets handpiece reduction ratio.  See the available reduction ratio in the following table:  
 
1:1 
16:1 
 
:
RPM +/‐
 Sets motor rotation speed (value in rpm). Speed values depend on contra‐angle reduction rate.  
Max speed is 2000 rpm with 20:1 contrangle. Min speed is 20 rpm with 20:1 contrangle. 
Ncm +/‐: increases or decreases the torque values. Torque values are in Ncm and depend on contra‐angle reduction rate.  
Max torque is 80 Ncm with 20:1 contrangle. 
Pay  special  attention  to  use  the  same  ratio  of  the  handpiece.  Torque  accuracy  is  guarantee  only  with  20:1  Contrangle  correctly 
maintained. The torque values are only for handpieces that operate properly. Micro motor stops when set torque is reached. The torque 
value in counterclockwise rotation is always = Max Ncm 
LED = LED ON/OFF 
:
FLOW
 Selects variable irrigation pump flow rate. See the following table: 
 
0 = no irrigation 
 
:
F/R
 Switch between forward and reverse motor rotation (when in reverse, console will emit a beeping tone).
:
MOTOR ON/OFF
 Switches motor ON and OFF (if enabled). PUMP ON/OFF
 
 
USER MANUAL Rev 25/05/2023 
 
20:1 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
I TECH LITE 
Page 9 of 16 
 
9.Use 
 
 
24:1 
32:1 
:
 Switches pump ON and OFF (if enabled). 
 
 
 
64:1 
80:1