After the unit has been set up and the user has become familiar with the System's control panel functions:
1.
Attach the appropriate "E‐Type" handpiece to the motor. It can be used with most surgical contra‐angles without internal
light or spray on the market. Use only handpiece that conform to ISO 3964. Change the handpiece and angle piece only when
the motor not running.
1.
Insert bag hanger rod into socket on the top of the console.
2.
Attach the motor to the connector on the right side (facing the unit) on the front panel.
3.
Attach the foot control to the connector (gray color) on the front panel.
4.
Install irrigation tubing into pump on the rear panel. Use only tubing as described in "Irrigation Tube Set Up".
5.
Switch on (on the rear panel.
In every program Programs [ 1 to 5] are set the same parameters by the manufacturer.
Perform torque calibration, press "Cal" key on front panel while holding the motor. This function checks the handpiece mechanical
inertia, to off‐set this value.
This operation will take a few seconds, the drill will run at different speed.
The drill must not be engaged in any operation when this function is performed.
Calibration should be done before each surgical session or anytime the handpiece is changed or lubricated.
I TECH LITE will store the operative parameters for each "program" as soon as the user sets them. They will be maintained in the memory
even if the central unit is switched off. For each "program" it is possible to set speed, torque, ratio, direction of rotation and flow. To
reset the parameters of the manufacturer (SystemReset), press "RESET" key on Display screen N° 2. Wait until the end of the operation.
9.1Programming the unit
:
R/R
Sets handpiece reduction ratio. See the available reduction ratio in the following table:
1:1
16:1
:
RPM +/‐
Sets motor rotation speed (value in rpm). Speed values depend on contra‐angle reduction rate.
Max speed is 2000 rpm with 20:1 contrangle. Min speed is 20 rpm with 20:1 contrangle.
Ncm +/‐: increases or decreases the torque values. Torque values are in Ncm and depend on contra‐angle reduction rate.
Max torque is 80 Ncm with 20:1 contrangle.
Pay special attention to use the same ratio of the handpiece. Torque accuracy is guarantee only with 20:1 Contrangle correctly
maintained. The torque values are only for handpieces that operate properly. Micro motor stops when set torque is reached. The torque
value in counterclockwise rotation is always = Max Ncm
LED = LED ON/OFF
:
FLOW
Selects variable irrigation pump flow rate. See the following table:
0 = no irrigation
:
F/R
Switch between forward and reverse motor rotation (when in reverse, console will emit a beeping tone).
:
MOTOR ON/OFF
Switches motor ON and OFF (if enabled). PUMP ON/OFF
USER MANUAL Rev 25/05/2023
20:1
1
2
3
4
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]
I TECH LITE
Page 9 of 16
9.Use
24:1
32:1
:
Switches pump ON and OFF (if enabled).
64:1
80:1