Auber Instruments AT210 Manuale di istruzioni - Pagina 2

Sfoglia online o scarica il pdf Manuale di istruzioni per Termometro Auber Instruments AT210. Auber Instruments AT210 3. Digital
Anche per Auber Instruments AT210: Note sull'uso (2 pagine), Manuale di istruzioni (3 pagine), Manuale di istruzioni (4 pagine)

AUBER INSTRUMENTS
AH1 and AH2 are called alarm on temperature, i.e. the temperatures at which the
Alarm1 and Alarm2 will turn on; AL1 and AL2 are called alarm off temperature, i.e. the
temperatures at which the Alarm1 and Alarm2 will turn off. When AH1>AL1(or
AH2>AL2), the alarms is set as a high limit alarm. When AH1<AL1(or AH2<AL2), the
alarm is set as a low limit al a rm. For a high limit alarm, the alarm buzzer shoul d be on
once the temperature ri s e above AH, and it should be off only when the temperature
drops below AL. In contrast, for a low limit alarm, the alarm buzzer shoul d be on when
the temperature drops below AH, and it should be off only when the temperature rise
above AL.
For example, when the temperature unit is set to F (Fahrenheit), AH1=900, and
AL1=800, the buzzer wil l go off once the measured temperature is higher than 900° F;
the buzzer wil l stop when the temperature drops below 800° F. In another case, when
AH2=180, AL2=185, the alarm2 is set as a low limi t alarm. If the temperature drops
below 180° F, the buzzer wil l go off (on); when the temperature rises above 185° F, the
buzzer will stop.
Users can press the Shift key ( > ) to mute the buzzer. The alarm will buzz again if the
alarm temperature moves out of the alarm zone (between AH and AL) and re-enters it
again. For i n stance, AH1=900, AL1=800, current temperature is higher than 900, and
the alarm is muted. In this case, the temperature has to drop below 800 and rise
above 900 again to trigger the alarm buzzer. To di s abl e the alarms, set AH and AL to
the same value, i.e., AH1=AL1 and AH2=AL2.
Steps to set an alarm are described below as well as in the flow chart in Figure 4:
(1) Press SET key, change the code to 0001, and press SET again to enter alarm
setting mode.
(2) Press Up and Down key to select a parameter (AH1, AL1, AH2, or AL2).
(3) Press SET to view the value of the parameter.
(4) Use Shift, Up, and Down key to change the val u e.
(5) Press SET key to confi r m the value and exit from this parameter.
(6) Press Up or Down key to select the new parameter.
(7) To exit this mode, press SET key when End is displayed.
The procedure is showed in the flow chart in Figure 4.
Parameter Display
Oper ation Mode
Enter Code
SET
000
SET
XXXX
1
SET
Figure 4. Steps of setting alarms. In this example, alarm1 is set as a high limit alarm at
480; alarm2 is disabled.
7. System settings (code 0089)
The parameters for system configuration is listed in Table 1.
Table 1. System configuration parameters.
Symbol
Name
Description
Int y
Input sensor type
Inty
PSb
PSb
Input of fset
FILT
Digital f ilter
FILT
C-F
C-F
Temperature unit
Brit
Brit
Brig htne ss level of dimme d LED
END
END
Note 1. Inty (Input sensor type). This parameter defines the sensor type that is used for
the thermometer. If you purchased the thermometer bundle, it is already set for the
probe included in the package. You don t need to change it.
Note 2. PSb (input offset). This is to set an offset value to compensate the error
produced by the sensor. For example, if the meter displays 5 ºC when probe is in ice/
water mixture, setting PSb= -5 will make the meter display 0 ºC. You can use shift key
(second key from the left) to shift to the first left digit, then press down key once to enter
minus sign.
Note 3. FILt (digital Filter). If the temperature reading fl u ctuates due to input noise, a
digital filter can be used to smooth the input signal. FILt may be configured in the
range of 0-3. Higher the value, stronger the fi l tering effect. A strong filter increases the
stabi l ity of the readout, but causes longer delay in responding to the temperature
change. Set FILt=0 will disable the filter.
Note 4. C-F (temperature unit). Set to C if you want to display temperature in Celsius.
Set to F for Fahrenheit.
2015.08
Buzzer On Temp
SET
SET
0480
AH1
Buzzer Off Temp
SET
047
SET
AL1
9
Buzzer On Temp
SET
SET
0250
AH2
Buzzer Off Temp
SET
SET
0250
AL2
END
Sett ing
Range
Initial
K
k
-1000 ~ 8000
0
0 ~ 3
0
° C, ° F
° F
1-4
1
Exit
Note 5. Brit (display di m mer). This parameter sets the brightness level of the dimmed
LED display. You can switch the LED brightness level between the dimmed and the
normal using UP key. However, the normal bri g htness l e vel of the LED display is equal to
level 4. So if Brit is set to 4, you won t see and change in the brightness of LED display
when pressing UP key.
Note 6. End. Exit the parameter setting mode.
Steps to set system parameters are described below as well as in the flow chart in Figure
5:
(1) Press SET key, change code to 0089, and press SET again to enter the system
confi g uration mode.
(2) Press Up or Down key to select the parameter to be changed.
(3) Press SET to view the value of the parameter.
(4) Press Shift, Up, and Down key to enter a new value.
(5) Press SET to confi r m.
(6) Press Up or Down to select the new parameter.
(7) To exit the menu, press SET when End is displayed.
Oper ation Mode
Enter Code
SET
0089
XXXX
SET
Figure 5. Flow chart of how to set system configuration parameters.
8. Wireless connection setting (code 0110)
The parameters for wireless connection can be accessed by code 0110. The parameters
are listed in Table 2.
Table 2. Wireless connection parameters.
Symb ol
Name
bAud
bAud
Id
Id
tdly *
Hand shake time
Tdl y
tACy
tACy *
Sampling rate
Note
End
End
1
Note 7. Baud (Baud rate for communi c ation). This should be left as default.
2
Note 8. Id (Unit ID). Each slave unit should be assigned with a unique Unit ID. The
3
default value is 1. The value of the Unit ID can range from 0 to 240. A master unit can
communi c ate with up to 10 slave units. For example, if you have two slave units, you can
4
set their Unit ID to 1 and 2 respectively. Then, when the master Unit ID is set to 1, it will
5
read the temperature of slave unit 1. When the master Unit ID is set to 2, it will read from
6
slave unit 2. Set two slave units to the same ID number will cause crashes.
Note 9. tdly (Hand shake ti m e). This is the time interval (unit i n seconds) after which the
master unit will attempt to read signal from the slave unit again. If the master unit
receives signal during this hand shake time, it will refresh and display the current reading
at the assigned cycle rate. Otherwise, it will display the previous reading for the rest of
the current hand shake cycle. If there is still no signal received during the entire hand
shake time, the master unit wil l display four dashes ( - - - - ). This setting can prevent the
master unit from frequently switching between temperature reading and four dashes. If
the master unit and the slave uni t is far away from each other or the signal is weak, user
should increase the hand shake time. Placing the master and slave unit too close to each
other may also result in no readi n g on the master unit.
Note 10. tACy (Sampling rate). This parameter sets how frequently the slave unit refresh
its reading . The default sampli n g rate is 1 per second. The unit is s
Note *. These two parameters are only available on the master unit.
WWW.AUBERINS.COM
Parameter Display
Input Type
SET
SET
xxxx
Inty
Input O ffset
SET
xxxx
PSb
Digital Filter
SET
xxxx
FiLT
Temper ature Unit
SET
xxxx
C-F
Br ightness level of dimmed
SET
xxxx
Brit
End
Description
Sett ing Range
12 00, 2400, 3600,
Bau d rate
9600
Unit ID
0 ~ 240
2 ~ 60
0.5 ~ 60 per second
Exit
-1
LED
Initial
Note
96 00
7
1
8
3.0
9 *
1.0
10 *
or Hz.
P2/3