9. Peak value setting (code 0037)
The peak value (the highest measured temperature) and the peak time (the time at
which the peak values i s read) can be stored. These parameters for peak values can
be accessed by code 0037. The parameters are listed in Table 3 below.
Table 3. Alarm setting parameters.
Symbol
Name
Description
mA
mA
Peak value
mAt
Peak value time
mAt
End
End
Exit
Note 11. mA (Peak Value). The highest temperature reading. By defaul t this function is
turned on. The Peak Value will be kept in the memory unless cleared. After the unit is
powered on, the previously recorded Peak Value and its time stamp will remain in the
memory unless: 1) a new peak value overwrites the previous peak val u e, 2) or the
memory has been reset. Hold the Down key for 2 seconds will reset both the peak
value and its time stamp to 0. The recorded peak value can be checked by pressing
DOWN key during the normal operating mode; press DOWN key again to display the
Peak Value Ti m e; press and hold Down key for 2 seconds to clear the stored Peak
Value and the Peak Value Time.
Note 12. mAt (Peak Value Ti m e). The time stamp of the Peak Value. By defaul t this
function is turned on. The time is tracked by the build-in timer which is activated
whenever the mAt function is bei n g turned on. The current Peak Value Time wi l l be
overwritten by the new Peak value time. The Peak value time can be checked by
pressing DOWN key again in the peak value mode. Pressing the DOWN key 2 seconds
will clear the Peak value as well as the Peak value time. Every time the unit is restarted
or the Peak Value is cleared, the timer will start over agai n (unit i n seconds).
12. Communication Between the Master Unit and Slave Units.
When the communication between a master unit and a slave uni t is establ i shed, all the
settings and operations on these two units are mirrored , except for the Unit ID and
the operation which temporarily mute the alarm (for the safety reason). You can set al l
of these parameters from either the slave or the master unit.
1) Mirrored operations and parameters. Operations and readings on one unit will be
mirrored to the other unit. Change of the alarm setting on the master uni t will also be
synchronized to the slave unit, or vice versa. The temperature reading and peak
values on the slave unit will also be synchronized between the master unit and the
slave unit. For example, when the peak value is cleared from the slave unit, the new
peak value will be recorded at the slave unit and then mirrored to the master unit.
When the peak val u e is cleared from the master unit, this operation will also be
synchronized to the slave unit, and the peak value will be cleared.
2) Muting the alarms. When you press the SHIFT key to mute the alarm buzzer, it will
only be effective to the buzzer on this unit. The buzzer on the other unit will continue
sound until temperature left the alarm zone, or the SHIFT key on that unit is pressed.
3) Placing distance. Placing the master and the slave units too far or too close will
result in bad signal reception on the master unit. Signal reception can be affect by
distance, physical obstacles, and other factors. The master unit and the slave unit can
be placed 100 feet away from each other in an open space.
Figure 6. Velcro mounti n g (left) and stainl e ss steel plate mounting (right).
2015.08
Sett ing Range
Initial
Note
On/off
On
11
On/off
On
12
11. Placing and Mounting the Meter.
The meter should be pl a ced in a working environment where temperature is -20-50° C
(0-100° F). It should be pl a ced away from hi g h heat to protect the plastic housing and
electronics. Two mounting opti o ns are provided:
1) A pair of Velcro fastener. The Velcro has a pressure sensi t ive adhesive backing. You
can remove the protective fi l m from the hook piece and stick it to the back of the meter
(see Figure 6). Then remove the protective film from the loop piece and install it onto
the wall. Please note that the pressure sensitive adhesive on Vel c ro is industrial grade
with strong holding force. It needs to be stick on a solid surface. Don t put it on a drywall
because i t may peel off the paint if you decided to remove it later.
2) A stainless steel mounting plate. The plate allows the meter to be hung on a hook or
a nail. It also allows the meter to be permanently mounted with a screw. The bottom
part of the stai n less steel mounti n g plate is covered with pressure sensitive adhesive.
To install it onto the meter, peel off the pink colored releasing film; press it firmly onto
the back of the meter (see figure 6).
NOTE: This equi p ment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference i n a residential installation.
This equi p ment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of the
following measures:
1) Reorient or relocate the receiving antenna.
2) Increase the separation between the equi p ment and receiver.
3) Connect the equipment into an outlet on a circuit different from that to which the
receiver i s connected.
4) Consult the dealer or an experienced radi o /TV technician for help.
Caution: Any changes or modifications to this device not explicitl y approved by
manufacturer could voi d your authority to operate this equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference, and (2)
this device must accept any interference recei v ed, including interference that may
cause undesired operation.
Auber Instruments
5755 North Poi n t Parkway, Suite 99
Alpharetta, GA 30022
www.auberins.com
Email: [email protected] Tel: 770-569-8420