9. Operate the Shutter by pressing the Footswitch. With the Shutter Selector Switch in the manual position, the
Shutter operates directly from the Footswitch. In the timed position, the Shutter opening is determined by the
setting on the electronic Timer. Simply push the Timer Setting Buttons to enter the desired number of
seconds the Shutter is to remain open.
BULB REPLACEMENT PROCEDURE
Bulb replacement is easily accomplished by following the steps
below. Refer to the labeled diagrams under the light source
cover.
1. Ensure that the Power Cord is unplugged from the rear of
the unit.
2. If the unit was previously running, allow the Bulb to cool 10-
15 minutes.
3. Remove the top cover from the BlueWave 50 by loosening
the four cover fasteners.
4. Disconnect the Bulb Plug (Figure 8).
5. Loosen the two Thumbscrews and lift the Bulb Retaining Clip
(Figure 8)
6. Remove the Bulb Reflector Assembly (Figure 9).
7. Install the new bulb by reversing steps 4 through 6.
8. Reinstall the unit's cover and plug in the Power Cord.
9. Turn the Power Switch on. The Bulb will now ignite. Allow
the Bulb to warm up for 5 minutes. The unit is ready for use
once the Bulb has warmed up.
®
BlueWave
50 Light-Curing Spot Lamp
Bulb Plug
Bulb
Retaining
Clip
Bulb Reflector Assembly
Figure 7. Bulb Assembly Components
Figure 8. Bulb Removal
Figure 9. Bulb Reflector Removal
Thumbscrews
11