AUTO PULSE MODE
Placing the device into Auto Pulse Mode will initiate the automatic delivery of pulse doses so oxygen purity
can be checked. This mode is only accessable to the provider by pressing a combination of buttons on the
Control Panel. follow the steps below to check the device's purity level.
Recommended Oxygen Purity Measurement Device: PRO2 Check Elite Multifunction Oxygen Indicator by Salter Labs
1. Power on the device.
2. Press and release the "OK", right arrow "
will display the following:
3. Select the desired Pulse Setting.
NOTE: This mode delivers a bolus of oxygen at 15 bPM across all Pulse Settings (1, 2, 3 or 4).
4. Connect 1/8" silicon tubing to the oxygen flow port of the measurement device and the Cannula Nozzle
fitting on the POC.
5. Allow approximately five minutes for the device to reach full oxygen concentration and to stabilize. if you
change the Pulse Setting, allow the device time to stabilize at the new setting before recording readings.
NOTE: it is normal for oxygen concentration values to cycle up and down. The highest value should be
considered final for the device. if the swing is more than 2%, the unit is not yet stable enough to measure.
6. To exit Auto Pulse Mode, press the "OK" button or hold down the "Power" button until the unit turns off.
when powered back on, the device will be in normal operating mode.
ELECTRONIC RESET
if the display screen does not appear as described in the Operator's Manual or if the Reset Required alarm
is activated, follow the instructions below to reset the device to default factory settings:
1. Power the device off.
2. After the device has turned off, press and hold the "Power" button until the display screen stops flashing
and the display turns blank. Once the display is blank, release the "Power" button.
3. Allow appoximately 15 seconds for the device to cycle through a short sequence. during this time, the
device will power on and off.
4. Power the device back on. Verify that the display screen is appearing as described in the Operator's
Manual. if this reset does not resolve the issue, contact inova Labs Customer Care at 1.512.617.1744 or
your inova Labs Sales Representative.
+