1. Does the Activox 4L model have scheduled maintenance?
Like all POCs, the Activox 4L will eventually require a sieve bed change. in the Service Life section of the
Operator's Manual, it is suggested that the patient run the device weekly for a few hours if it is has not
been used on a regular or daily basis. frequent Low Oxygen Purity and high Temperature alarms may
indicate that the system requires service.
2. What type of rechargeable batteries are in the LifeChoice Activox 4L and what are the
recommendations to extend the overall service life of the rechargeable batteries?
All inova Labs POCs use Lithium-ion rechargeable batteries. The characteristics of Lithium-ion
rechargeable batteries include: high voltage/light weight, no memory effect, and flat discharge voltage.
• A partial discharge reduces stress and prolongs battery life. it is recommended the battery be
recharged, if possible, prior to total discharge. The Operator's Manual informs the patient to "recharge
the battery, even if only partially depleted as often as possible.
• Extreme temperature swings effect the battery's cycle life. Leaving the device or external battery in
a hot car, especially if exposed to the sun, is very stressful and can damage the cells.
• when not in use, store the device and/or External battery is a cool, dry place. for long-term storage,
we recommend keeping the device on a charger full-time so the it will retain sufficient charge to
perform Repressurization. This will help extend sieve bed life. for long- term storage of the External
battery, we recommend keeping about a 40% charge on the battery. This allows for some self-
discharge while still retaining sufficient charge to keep the protection circuit active.
3. What is the standard warranty of the LifeChoice Activox 4L?
The basic description of the warranty is that from date of shipment from inova Labs, the device's internal
hardware is covered for three years while the batteries, sieve bed and accessories are covered for one
year. Please refer to the Limited warranty that is included in the Sales Agreement and in every Operator's
Manual for a complete description.
4. How do I initiate a return for service to Inova Labs?
Prior to requesting a return, please troubleshoot over the phone with inova Labs Customer Care Team at
1.512.617.1744 or your Territory Manager. you may request an RMA by phone to Customer Care or by
email to [email protected]. include the serial number of the unit, a detailed description of the
problem, your contact information and the return shipment address.