AFJ Instruments LT32C/10 Manuale operativo - Pagina 18

Sfoglia online o scarica il pdf Manuale operativo per Hardware di rete AFJ Instruments LT32C/10. AFJ Instruments LT32C/10 18. Artificial mains network (lisn)

SECTION G: LT32C/10 REMOTE CONTROL
G.1 • GENERAL
To use AFJ LISN LT32C/10 in remote mode, input signal REM
signal (visualized by LED) became active
The signal S0 and S1 coincidence define the line selection as per the following list (Table G.1):
REM
in
H
L
L
On the remote control connector there are four signals (Figure G.1):
1. S0 input signal;
2. S1 input signal;
3. REM
input signal;
in
4. REM
Output signal (open collector).
out
To make INPUT SIGNALS active is shall be sufficient putting them to ground.
AFJ LT32C/10 LISN - Operating Manual - rev. 3.8 / 2019-01
S0
S1
X
X
H
H
L
H
Table G.1: LT32C/10 signals combination
Figure G.1: Remote control connector
Distributed by:
Reliant EMC LLC
3311 Lewis Ave, Signal Hill CA 90755,
408-916-5750, www.reliantemc.com
has to be grounded, so that REM
in
REM
Active line
out
H (Led=off)
Manual
L
L
out
N
L
18