Game ready ATX Manuale d'uso
Sfoglia online o scarica il pdf Manuale d'uso per Accessori per videogiochi Game ready ATX. Game ready ATX 6. Full leg boot for use with grpro 2.1 system and med4 elite system
FOR USE WITH
GAME READY
GRPRO
2.1 SYSTEM
®
MED4 ELITE™ SYSTEM
SIZING SPECIFICATIONS
Actual Wrap Length
Size
(Adjustment ranges)
M
30–32"
(Medium)
L
34.5–36.5"
(Large)
XL
39–41"
(Extra Large)
BEST PRACTICES
•
Keep the Control Unit (GRPro 2.1) on a table to prevent drainage
from the Wrap.
•
If you are using two Full Leg Boot wraps (one on each leg), we
recommend using the Med4 Elite or two GRPro 2.1 Control Units to
administer therapy. We do not suggest using the dual connector hose
with the GRPro 2.1.
•
Never cover the toes when
applying the Full Leg Boot.
PRIMING THE WRAP
To prevent 04 errors at the beginning of therapy with the Full Leg Boot
when used with the GRPro 2.1 Control Unit, it is important to prime
(pre-fill) the Wrap before a therapy.
To prime the Wrap for use with the GRPro 2.1 Control Unit:
1. With the GRPro 2.1 Control Unit off, attach the Connector Hose to
the Control Unit and the Wrap.
2. Lay the Wrap open and flat next to the Control Unit (not on the
body).
3. Turn the System on and run a therapy with no compression. Pause
after 45 seconds.
4. Disconnect the Full Leg Boot from the Control Unit so water
remains in the Wrap. It is now ready for use.
Approximate
Anatomical
In-Seam
Orientation
27–33"
All sizes
33–37"
fit left or
right leg
37"+
FULL LEG
WARNING
It is mandatory to fully read and understand your
System's User Manual before using the device.
Failure to follow operating instructions could result in
serious injury.
IMPORTANT
Read complete indications, contraindications,
cautions, and warning before using this product.
Keep this document for future reference.
NOTE: If you are priming the Full Leg Boot for the first time
using the GRPro 2.1, you may encounter a 04 Error 60 seconds
after pressing play. If this occurs, power off the Control Unit by
pressing the power button. Power it back on, and repeat the
steps listed above.
After completing the previous steps, check the water level in the Control
Unit. If necessary, add more water. Be careful not to overfill the Control
Unit, and allow room for water to return from the Wrap after therapy. Check
the User Manual of your Control Unit for more information on adding water.
If you do not plan to use the Full Leg Boot within the next 24 - 48 hours,
allow the water to drain from the Wrap back into the Control Unit by
leaving it connected to the Control Unit for 2 to 3 minutes. This will
help prevent mildew from forming inside the Wrap and keep your Wrap
performing as expected.
NOTE: It is not necessary to prime the wrap for use with the
Med4 Elite Control Unit.
If water has been in the Wrap for more than one hour, we recommend
pre-cooling/pre-heating the Wrap prior to running a therapy on a patient.
See Pre-Cooling/Pre-Heating Your Wrap for instructions on how to pre-cool/
pre-heat the Full Leg Boot.
BOOT
USE GUIDE
ENGLISH