Janam GT2 Series Manuale di calibrazione - Pagina 2
Sfoglia online o scarica il pdf Manuale di calibrazione per Regolatore di temperatura Janam GT2 Series. Janam GT2 Series 3.
Anche per Janam GT2 Series: Manuale di avvio rapido (12 pagine)
Instructions
Step 1
Using an FDA-approved thermal forehead thermometer, take ten temperature measurements of a
single test subject (person). Record the average value of all ten measurements. (To find the average
value, add all ten measurements and divide by ten. This final number will be your average value.)
Example:
1 PERSON
10
X
Step 2
Using GT2's temperature reading functionality, take ten temperature measurements of the same
single test subject used in Step 1. Record the average value of all ten measurements.
Example:
1 PERSON
10
X
Janam Technologies LLC
Measurement #
Value
1
98.6
2
98.7
3
98.6
...
...
10
98.6
Total = 986 (divide by 10)
Average Value = 98.6
Measurement #
Value
1
98.5
2
98.6
3
98.7
...
...
10
98.7
Total = 985 (divide by 10)
Average Value = 98.5
GT2
Calibration Guide
1