Janam GT2 Series Manuale di calibrazione - Pagina 3

Sfoglia online o scarica il pdf Manuale di calibrazione per Regolatore di temperatura Janam GT2 Series. Janam GT2 Series 3.
Anche per Janam GT2 Series: Manuale di avvio rapido (12 pagine)

Step 3
Compare the average values from Steps 1 and 2. If these two values are different, the temperature
reading must be calibrated. To prepare for calibration, complete the following steps:
1. Connect to Wi-Fi, record the IP address on the screen
2. Input the IP address plus +port 8090 in the PC Chrome window (Example: 192.168.8.2∶8090)
3. Access the control panel by entering the password "123456"
4. Access
Settings
by clicking the gear icon on the blue menu bar at the bottom of the menu
Step 4
In the temperature measurement calibration field, enter the difference between the FDA-approved
thermometer (Step 1) and GT2's temperature reading functionality (Step 2). For the example used,
the value that should be entered is "0.1".
Positive calibration ("compensation") is between 0.0 ~ 10.0
Negative calibration ("compensation") is between -0.0 ~ -10.0
Janam Technologies LLC
GT2
Calibration Guide