A+ Corporation, LLC. Genie Direct Drive 752 Istruzioni per l'installazione e il funzionamento - Pagina 4
Sfoglia online o scarica il pdf Istruzioni per l'installazione e il funzionamento per Apparecchiature di prova A+ Corporation, LLC. Genie Direct Drive 752. A+ Corporation, LLC. Genie Direct Drive 752 5.
An ISO 9001:2008 certified company
Removing Probe for Service
Step 1. Retract the probe
Retract the probe by using the 7/16" open end wrench only on the wrench flats of the threaded probe.
}
Turn the threaded probe until it comes to a noticeable stop in its original position. (See Figure 2)
}
Step 2. Isolate and Vent the Probe
Slowly close the ball valve to the process. If resistance is meet when closing the valve be sure the probe is retracted 100%.
}
Note: If the probe is closed in the ball valve, the threads on the probe could be damaged.
Use the factory supplied vent valve in the 1/8" NPT auxiliary port and a 7/16" wrench to bleed remaining process that is trapped between the
}
closed process valve and the probe body. (see Figure 1)
Step 3. Remove from process valve
Remove the probe from the process valve using a 1-3/8" open end wrench ONLY on the lowest wrench flats on the base, nearest to the NPT
}
male threads (see Figure 1)
Figure 1
Figure 2
Vent
Valve
Figure 3
D1
4
Figure 4
Figure 5
D1
©
2010 A+ Corporation, LLC. All rights reserved.
SCC-752-InsInstallationOperation_0514