A.H. Systems EHA-51B Manuale operativo - Pagina 11

Sfoglia online o scarica il pdf Manuale operativo per Antenna A.H. Systems EHA-51B. A.H. Systems EHA-51B 14. Active monopole and loop antenna

A.H. Systems EHA-51B Manuale operativo
A.H. Systems Model EHA-51B Active Monopole and Loop Antenna

CALCULATIONS

EMISSIONS TESTING
Individual calibration data for the active loop antenna is supplied to comply
with various emissions test requirements. For emissions measurements, add
antenna factor plus cable loss to receiver reading in dBV to convert to field
strength in dBA/meter.
Field Strength(dBuA/m) = SA(dBuV) + MAF(dB/m) + cable loss (dB)
Fictitious electric field strength can be calculated by the following:
 A.H. Systems inc. – Oct 2017
REV B
SA = Spectrum Analyzer or Receiver voltage reading
MAF = Magnetic Antenna Correction Factor
CL = Cable Loss in dB
dBV/m = dBA/m + 51.5
11