A.H. Systems EHA-51B Bedieningshandleiding - Pagina 11
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} A.H. Systems EHA-51B. A.H. Systems EHA-51B 14 pagina's. Active monopole and loop antenna
A.H. Systems Model EHA-51B Active Monopole and Loop Antenna
CALCULATIONS
EMISSIONS TESTING
Individual calibration data for the active loop antenna is supplied to comply
with various emissions test requirements. For emissions measurements, add
antenna factor plus cable loss to receiver reading in dBV to convert to field
strength in dBA/meter.
Field Strength(dBuA/m) = SA(dBuV) + MAF(dB/m) + cable loss (dB)
Fictitious electric field strength can be calculated by the following:
A.H. Systems inc. – Oct 2017
REV B
SA = Spectrum Analyzer or Receiver voltage reading
MAF = Magnetic Antenna Correction Factor
CL = Cable Loss in dB
dBV/m = dBA/m + 51.5
11